EasyManua.ls Logo

Accu-Chek Nano - Understanding Out-of-Range Control Test Results

Accu-Chek Nano
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
5. Remove the control bottle cap and wipe the tip of
the bottle with a tissue.
6. Squeeze the bottle until a tiny drop forms at the
tip. Touch the drop to the front edge of the yellow
window of the test strip. Do not put control
solution on top of the test strip. When you see
flash, you have enough control solution in the
test strip.