Performance Analysis
The performance data for the Accu-Chek Performa Nano System (Accu-Chek
Performa Nano Meter with Accu-Chek Performa Test Strips) were obtained using
capillary blood from diabetic patients (method comparison, accuracy), venous blood
(repeatability), and control solution (reproducibility). The system is calibrated with
venous blood containing various levels of glucose. The reference values are obtained
using the hexokinase method. For method comparison, the results were compared
with results obtained using the hexokinase method with deproteinization (automatic
analyzer). The hexokinase method is traceable to an NIST standard.
Measuring Principle
Refer to your test strip package insert for more information.
75
43149_NanoPerf_24MSTR_Ins.indd 7543149_NanoPerf_24MSTR_Ins.indd 75 11/25/08 2:49:10 PM11/25/08 2:49:10 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Performa Nano and is the answer not in the manual?

Accu-Chek Performa Nano Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Battery LifeApproximately 1000 tests
Dimensions69 x 43 x 20 mm
DisplayLCD
Memory Capacity500 results
Automatic Power Off90 seconds
Battery1 x 3V lithium battery (CR2032)
Test StripAccu-Chek Performa
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Measurement Range0.6 - 33.3 mmol/L
WeightApprox. 40 g (with battery)
Operating Conditions6°C to 44°C, 10% to 90% RH