Meter Memory, Setup, and Data Transfer
3
Memory
Storing Blood Glucose and Control Results
The meter automatically stores up to 500blood glucose results and up to 20 control results with the time
and date of the result and any test result marker which you can review at any time. Test results are stored
from the newest to the oldest, so set the time and date correctly in the meter. Having the correct time and
date setting helps ensure appropriate interpretation of stored blood glucose results by you and your
healthcare team.
NOTE
The memory is not lost when the battery is replaced; however, confirm that the time and date are still
correct. See Chapter3, Setting the Time and Date.
Once 500blood glucose results are in memory, adding a new test result causes the oldest one to be
deleted.
If more than 500blood glucose tests are performed within a 90-day period, only the most recent
500test results are included in the 90-day average.
Control results are stored in memory, but cannot be reviewed on the meter. To view stored control
results, first transfer them to a compatible software application. Contact Roche for product
availability.
Control results are not included in the 7-, 14-, 30-, and 90-day averages.
Only test results that have been assigned a pre-meal or post-meal marker are included in pre-meal
and post-meal averages. All blood glucose results are included in the general 7-, 14-, 30-, and
90-day averages.
Chapter 3: Meter Memory, Setup, and Data Transfer
51746_07022590001.indb 17 7/11/13 1:27 PM

Table of Contents

Other manuals for Accu-Chek Performa

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Performa and is the answer not in the manual?

Accu-Chek Performa Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Meter Dimensions94 x 52 x 21 mm
Measurement Range0.6-33.3 mmol/L
Memory Capacity500 results
Battery Type1 x CR2032 lithium coin cell
Automatic Power Off90 seconds after last user action
Operating Humidity10 to 90% RH
Storage Conditions10% to 90% RH
Meter Weight59 g (with battery)