34
Maintenance and Troubleshooting
5
Display Action
Blood glucose is below the defined hypoglycaemic (low blood glucose) level. See
Chapter2, Unusual Blood Glucose Results.
A general marker was assigned to this test result.
A pre-meal marker was assigned to this test result.
A post-meal marker was assigned to this test result.
A pre-meal marker was assigned to this test result and the post-meal test reminder
has been activated.
The test strip may be damaged or not properly inserted. Remove and reinsert the test
strip and replace it if damaged.
Your blood glucose may be extremely high or a meter or a test strip error has
occurred.
• If your test result matches how you feel, contact your healthcare professional
immediately.
• If your test result does not match how you feel, repeat the blood glucose test. See
Chapter2, Unusual Blood Glucose Results.
• If the E-3code still appears for your blood glucose test, your blood glucose
result may be extremely high and above the system’s reading range. Contact
your healthcare professional immediately.
• If the second test result does not match how you feel, perform a control test
with the control solution and a new test strip.
• If the control result is within the acceptable range, review the proper
testing procedure and repeat the blood glucose test with a new test strip.
• If the control result is not within the acceptable range, see Chapter4,
Understanding Out-of-Range Control Results.
51746_07022590001.indb 34 7/11/13 1:27 PM