6
1
5
4
5
3
2
7
4 manual 19-11-2013
2. SPECIFICATIONS OF THE FLOWFINDER MK2
®
SPECIFICATIONS FF MK2
Air ow range (supply and exhaust) 10..550 m³/uur with zero pressure compensaon
551..850 m³/uur with calculated compensaon
Air temperature range -20..+80
o
C
Uncertainty (@ 20°C) * 3% of the reading, with a minimum of 3 m³/hr
Repeatability beer than 1%, with a minimum of 3 m³/hr
Resoluon 1 m³/hr, 0.1cfm, 0.1l/s
RV range 0..100%
Full charge capacity 8 hours
Weight 2230 gram baery pack included
Dimensions FlowFinder
mk2
ø 260 x 350mm
Use the packing list to verify all
elements of the FlowFinder mk2® are
present. Avoid liing the FlowFinder mk2®
while holding onto a hood.
* Uncertainty is dened as the square root of the sum of the squares of non-lineairity, hysteresis and repeatability.
1. PARTS OF THE FLOWFINDER
MK2
1. FlowFinder mk2® with display
2. Transport case (with key)
3. Baery pack (mounted onto the
FlowFinder mk2®, or stored in the case)
4. Baery charger
5. Hood 230x230x145mm
6. Hood 600x600mm (oponal)
7. MicroSD card, USB adapter, SD adapter