5. Maintenance
5.1 Care and hygiene instructions
The unit cannot be sterilized.
All kind of wipe disinfectants that are listed below can be used for cleaning the unit. The complete
list with all active substances and product names is available in German on the Website of the “Ro-
bert Koch-Institut“ with the title “Liste der vom Robert Koch-Institut gepften und anerkannten
Desinfektionsmittel und –verfahren“ (source: Bundesgesundheitsblatt 10/2017) or from ACL.
The active substances in the following table are expected to be harmless for the unit as long as the
listed limits are observed. The table does not represent a medical evaluation of the effectiveness
of these substances.
Active substance Max. concentration in %
Max. exposure time in
minutes
Example products
Alcohol 100 360
Bacillol 30 Foam
Meliseptol
Biguanide 8 360 Incidin Plus
Chlorine, organic or
inorganic substances with
active chlorine
3 120 Trichlorol
Aldehyde and derivates 10 360 Optisept
Per-compounds 8 360
Incidin active
Terralin PAA
Phenol or phenolic
derivatives
6 360 Helipur
Make sure that the unit is cleaned and disinfected with an appropriate cleanser every time it is
sent back for repair. If you are not able to do so, mark the material as contaminated and use safety
foil to seal it twice.
To protect the health of our employees, we reserve ourselves the right to initiate suitable measu-
res on contaminated units at your costs or to refuse the unit at all.