Possible causes Solutions
The air pipe is crushed and leaks
Contact Acteon's Customer Service team to replace the
air pipe.
10.4 Powder coming out of the tank
Symptoms: powder is coming out of the tank
Possible causes Solutions
The tank plug is not screwed on properly. Screw the plug tightly onto the tank.
The plug O-ring is incorrectly positioned.
There are traces of powder under the plug O-ring.
l Remove the tank plug.
l Dispose of the powder in an infections clinical waste
container
l Remove the plug's O-ring and inspect it.
l Use the multi-purpose syringe's air function to blow
air and clean the lid threads.
l Place the O-ring back onto the plug.
l Fill the tank to the maximum level with suitable
powder
l Screw the plug back onto the tank.
The plug O-ring is defective.
Contact Acteon Customer Service team to replace the
plug seal.
The tank is cracked
Contact Acteon Customer Service team to replace the
tank.