11 Corrective Maintenance
In the event of faulty operation, the following corrective maintenance actions may be performed by the user.
11.1 Replacing the fuses
The medical device is protected by two fuses in the mains connector.
To replace the fuses, perform the following operations:
1. Stop the medical device (position O).
2. Disconnect the mains cord from the electrical network.
3. Disconnect the mains cord from the mains connector.
4. Insert the tip of a flathead screwdriver into the notch on top of the fuse holder to release it.
5. Remove the used fuses.
6. Replace the used fuses with fuses of the same type and same rating.
7. Place the fuse holder in its recess by pushing it until you hear a click that confirms it is in the correct position.
8. Connect the mains cord to the connector.
9. Connect the mains cord to the electrical network.
User Manual|AIRMAX|J11401|V7|(03)|12/2017|ND07EN010G - Page 29/42