OPERATOR’S MANUAL • X-Mind AC • (06) • 05/2018• NXACEN010F
CHAPTER 8
8.1. ERROR MESSAGES
An anomaly is indicated as follows:
- emission of an intermittent acoustic signal (beep)
- MALFUNCTIONING INDICATOR LED intermittently turns on
- the error code (E ….) appears on the display)
- all control panel functions are inhibited
In this case turn off the timer and then turn it back on. If the error persists, refer to the following table.
The following chart gives a list of error messages that may appear while “x-mind dc” radiographic system is
working.
The chart also includes the causes of the error messages to be communicated to the technical support and, in
some cases, what to do to solve them.
If the error is not xed by carrying out the described solution, turn off the power supply from the electrical panel
(wall mounting version) or unplug the system (mobile version)
ERROR
MESSAGES
CAUSE SOLUTION
E00
RX1 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E01
RX2 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
E02 CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
E03 EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
E07
LINE VOLTAGE VALUE NOT INCLUDED
WITHIN THE NOMINAL VALUE
CALL THE “ASSISTANCE SERVICE”
E08
THE X-ray KEY
ALWAYS SEEMS TO BE PRESSED
MAKE SURE IT IS NOT JAMMED
E09 ANOMALY IN THE CONTROL PANEL CALL THE “ASSISTANCE SERVICE”
E12
THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
KEEP THE X-ray KEY
PRESSED TILL THE END OF THE EXPOSURE
E20 ANOMALY IN THE TRIAC/RELAY CALL THE “ASSISTANCE SERVICE”
E21 ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE “ASSISTANCE SERVICE”
E22 ANOMALY IN THE CONTROL CIRCUIT CALL THE “ASSISTANCE SERVICE”
E23 INCORRECT DIP-SWITCH CONFIGURATION CALL THE “ASSISTANCE SERVICE”
E24
THE CONTROL BUTTON DOES NOT CORRESPOND
TO THE SELECTED TUBEHEAD
CALL THE “ASSISTANCE SERVICE”
ERR MAJOR ERROR
ALL FUNCTIONS ARE DISABLED
CALL THE “ASSISTANCE SERVICE”
88