3-4
12704A Resistivity Chamber Operation Manual
3.1.2 Connecting with the 8240
3.1.2 Connecting with the 8240
(1) Connect the INPUT terminal of the 8240 to the INPUT terminal of the 12704A by using
the TRIAX-TRIAX cable (A01009).
(2) Connect a voltage source such as the 6166 to the V SOURCE terminal of the 12704A by
using the output cable (CC010005) and the banana adapter (A08531).
The maximum output voltage of the 6166 is 1200 V. Perform measurement with
great care
The setting voltage must not exceed the maximum input voltage of the 8240.
If a cable is disconnected with voltage being applied from the V SOURCE, voltage
generation will occur on the end of the cable. Be sure to set the voltage STANDBY
before disconnecting the cable.
3.1.3 Connecting with the 8252
(1) Connect the INPUT terminal of the 8252 to the INPUT terminal of the 12704A by using
the TRIAX-TRIAX cable (A01009).
(2) Connect the HI OUTPUT terminal of the 8252 to the V SOURCE HI terminal of the
12704A by using the output cable (A01044).
(3) Connect the LO terminal of the 8252 to the LO terminal of the 12704A by using the
output cable (A01044).
(4) When using the LID SIGNAL function of the 12704A, connect the INTERLOCK on the
rear panel of the 8252 to the LID SIGNAL connector of the 12704A by using the
BNC-BNC cable (A01036). The INTERLOCK function of the 8252 operates differently
depending on the setting. For more information on the function and its setting method,
refer to the 8252 Operation Manual.
(5) Connect the LO terminal of the 8252 to the GND terminal.
WARNING