Anoxomat® III User’sGuide
Appendix A
Troubleshooting table
Note:
This symbol indicates a failure.
Error Cause Solution
I. Jar Connection Failure
Jar is not connected
properly (connection
is not used)
No jar connected Reconnect the jar to the jar tubing
The Anoxomat has detected
a vacuum condition in the
jar. The jar has been partly
processed by the Anoxomat
No jar connected
Release the vacuum in the jar by pushing into the
snap-shut coupling with an object (e.g., pencil)
Before reconnecting the jar, pull the outer ring of the
snap-shut coupling downward
II. Gas Connection Failure
Gas connection
has not passed the
corresponding test
No pressure from gas cylinder Check if gas cylinder is open
Pressure supplied to Anoxomat
should be 1.75 ± 0.15 bar (25.5±2.0
psi)
Check output pressure on the regulator
The regulator gauge does not
indicate actual pressure
Check actual pressure in the Process Report and adapt
the setting of the regulator
Gas cylinder is almost empty and can
no longer provide stable pressure
Check cylinder pressure on the regulator
Gas flow is restricted (e.g., a kink in
the gas tube)
Check the gas tube

Other manuals for Advanced Instruments Anoxomat III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments Anoxomat III and is the answer not in the manual?

Advanced Instruments Anoxomat III Specifications

General IconGeneral
BrandAdvanced Instruments
ModelAnoxomat III
CategoryMedical Equipment
LanguageEnglish

Related product manuals