Performing Unscheduled Volume Measurements
14
To measure a volume dis pens ed by a device without requ iring a calibration plan, use the Meas ure Volume option from the ! Home screen. Measurements
made with this option do not get recorded in the system database, no report is generated, and results are discarded after ending the measurement cycle.
1
2
3
From the ! Home screen Shortcuts bar,
click Calibrate Pipette. Choose PCS Volume
Quick Check for the Calibration Type.
Enter the expected target volume in µL, the
number of replicates to dispense, and select to
display errors as relative or absolute.
Select the instrument to use, click OK and then
follow the on-screen instructions for
completing the volume measurement session.
Ad hoc Calibration (alternative method)
Choose this Calibration Type to use a Calibration
Plan without affecting due dates for the pipette.
However, the calibration data will be saved in the
database for future reference.
*