Calibrating the PCS Instrument
From the ! Home screen, click Due
instruments if the count is not zero and
then skip to step 3. Otherwise, click the
Shortcut for Calibrate Instrument.
If calibrating from the Shortcut, select the
instrument from the instrument list and a
schedule if that is the calibration type.
Click Start Calibration.
Follow the on-screen instructions to
scan and insert PCS Instrument
Calibrator Kit vials A through D.
ArtelWare requires a Passing instrument calibration before performing pipette calibrations or operator trainings.
Artel recommends creating an Instrument Calibration schedule for every 30 days.
Note: The first time calibrating the PCS Instrument, ArtelWare requests the ArtelWare Activation code located inside the PCS Calibrator kit box. To use the Activation code before calibrating the
instrument, click ‘Scan Barcode’ on the ArtelWare ! Home screen and then follow the prompts in ArtelWare. This activates ArtelWare until the expiration of the date of the Calibration kit.
1
2
3
9
1
2
3
4
4
If calibrating from the Shortcut, select the
Calibration Type and click OK.