The Advanced
®
Osmometer Model 3320 Service Manual
The Event Record displays the last actions of
the instrument, and is used when servicing the
instrument. The output should look like:
Event Record output consists of:
# A sequential row number from
1 to 100.
Code: A number used to identify the
table containing the data defi-
nition.
Description: A text field describing the code
table.
Data: A numerical descriptor that,
when located in the appropriate
code table (see below), identi-
fies the action recorded.
Date/Time: The date and time the event
was recorded.
The code, data, and error numbers presented
in the Event Record can be decoded, using
the following tables.
Event Record
# Code Description Data Date/Time
===========================================
1: 0 State Change 2 [10/31/2006 11:02:11 AM ]
2: 0 State Change 9 [10/31/2006 11:02:14 AM ]
3: 4 Diagnostic Test 5 [10/31/2006 11:02:26 AM ]
4: 2 Bar Code Scan 0 [10/31/2006 11:02:33 AM ]
5: 5 Error 212 [10/31/2006 11:02:35 AM ]
6: 5 Error 207 [10/31/2006 11:33:51 AM ]
7: 0 State Change 7 [10/31/2006 11:33:53 AM ]
8: 8 Utility Routine 0 [10/31/2006 11:33:55 AM ]
9: 8 Utility Routine 2 [10/31/2006 11:33:57 AM ]
10: 5 Error 207 [10/31/2006 11:33:58 AM ]
End Event Record
Download Complete
INSTRUMENT STATES CODE: 0
DATA DESCRIPTION
0 RESET
1 ERROR
2 READY
3 STANDBY
4 TEST
5 CALIB
6 CALI_TEST
7 UTIL
8 SETUP
9 DIAG
10 BINTEST
SWITCHES CODE: 1
DATA DESCRIPTION
0 NONE
1 RIGHT
2 LEFT
4 TEST
8 SUPERVISOR
NOTE: Value reported is a combination of
above switches.
BAR CODE SCANS CODE: 2
DATA DESCRIPTION
NOTE: No subcategories.