The Advanced
®
Osmometer Model 3320 Service Manual
DIAGNOSTIC ROUTINES CODE: 4
DATA DESCRIPTION
0 A/D TEST
1 PROBE BIN TEST
2 SOLENOID TEST
3 DISPLAY/PRINTER TEST
4 KEYPAD/BEEPER TEST
5 BAR CODE TEST
SERIAL COMMANDS (instrument must be
in factory mode) CODE: 3
DATA DESCRIPTION
0 ? Help
1 ~ Reset Instrument
2 = Instrument Status
3 B Beep
4 C Add Comment to File
5 D Toggle Curve Trace On/Off
6 E Edit Parameters
7 H Block Temp (H+xxxxx mC)
8 I Impact (Ixx mS)
9 L Change Language
10 M A/D Channel Utilities
11 P Print Test
12 S Set Date/Time
13 T Show Date/Time
14 # Clear Parameters
15 1 Parameter Contents
16 2 Dump Results
17 3 Dump Events
18 4 Statistics
19 5 Manual Calibration
20 6 Memory Contents
21 0 Burn Parameters
22 Extra Carriage Returns
PARAMETER EDIT COMMANDS (instru-
ment must be in factory mode), subset of
code-3 data-6 CODE: 3
DATA DESCRIPTION
Solenoid Freeze Parameters
0 A On time
1 B Off Time
2 C Impacts
Solenoid Buzz Parameters
3 D On Time
4 B Off Time
5 C Impacts
Temperatures
6 G Supercool
7 H Post-Freeze
8 I TE Cooler Off
9 J Freeze Difference
10 K Buzz Difference
11 L Coldest
Calibration
12 M Slope
13 N Intercept
14 Y High Pt. Coefficient
15 O No. of Calibration Samples
16 P Reference Solution A
17 Q Reference Solution B
18 Z Reference Solution C
Probes
19 R Block Bin
20 S Sample Bin
Misc.
21 U Plateau Mode
22 V Serial Number
NOTE: The shared “CODE:3” is a bug,
recorded as #123 on 11/6/2006. Care
should be taken when decoding any serial
port logs after a code-3 data-6 sequence.