19
To ensure consistent and accurate pencil mark recogni-
tion your Advantage scanner should be calibrated annually or
after every 7,500 sheets processed, which ever comes rst. A
calibration sheet has been provided with your scanner for this
purpose. Please keep the calibration sheet in a convenient
location for easy retrieval as needed.
To determine how many sheets have been processed
since last calibration, press the Summary Display button 3
times. The display window will show how many sheets have
been scanned since the previous calibration.
If 7,500 sheets have been processed since your last
calibration the following message will appear on your display
window whenever the scanner is powered on. This message
will no longer display after the scanner has been calibrated.
READY: RECOMMEND
CALIBRATION SOON
Continued on page 20
Calibrating Advantage