Physical appearance check-after each cleaning and disinfection for the meter
Check Item
Losing display transparency
Losing legibility of any marking
Polymer crazing
Cracking
Swelling
Dissolving
Softening
Becoming brittle
Deformation
Acceptance Standard
Accept the result if none of the items
listed on the left appears after each
cleaning or disinfection process
Disinfection Process
After each cleaning or disinfecting process, you should
perform the physical appearance check and performance
check of the meter and lancing device for items shown in
the tables blow.
Step 4: The user should wash hands thoroughly with
soap and water after handling the meter, lancing device,
or test strips.
37