Advocate Redi-Code+ BMB-BA006A Blood Glucose Test Strips
Analysis Method Electrochemical
Physical Characteristics
Dimensions (mm) 30.9 x 6 0.58
Connector (pins) 6
Material
Base Material PET
Conductive Layer Ag & Carbon
Electrode Carbon
Chemical Components
Enzyme Glucose Oxidase
Mediator Electron Shuttle
Others Non-reactive ingredients
Function
Test Sample Whole Blood
Sample Size 1 µL
Test Time (sec.) 5 sec. / 7 sec.
Performance
Linear Range 20-600 mg/dL (1.1~33.3 mmol/L)
Precision CV < 5%
Accuracy <75 mg/dL ± 15mg/dL ; ≥ 75mg/dL ± 20%
Environmental Condition
Operation (Temp. / Humidity) 10°C - 40°C / RH below 85%
Storage (Temp. / Humidity) 4°C - 40°C / RH below 85%
Stability
Opened vial (days) 90 days
Closed vial (months) 24 months
Warning:
- Do not use expired strip
- This strip is only used with Advocate Redi Code+ Meters
- Close the can immediately after strip has been removed
- Fluoride: Cannot put uoride into blood as the preservative
- Iodoacetic Acid: It is impossible to put Iodoacetic Acid into blood as the preservative as well.
- New Born Baby: New born babies are not recommended to do this test.
- Reductant: High density of Vitamin C or uric acid may inuence glucose value
- Hematocrit: Hematocrit above 60% may cause lower glucose results. Hematocrit lower than 20% may cause higher
glucose results.
No Coding Required!
Test Strip
Specifications