Redi-Code+
No Coding Required!
Meter Specifications
Advocate Redi-Code+ BMB-EA001/BMB-EA001S Blood Glucose Monitor
Characteristics Specications
Dimensions (mm) 64 x 95 x 29
Weight (g) 53 (without batteries)
Micro Controller Name and Specications MSP430FG439
Power Source (2) 1.5V AAA alkaline batteries
Battery Low Warning Function (Voltage) 2.75V±0.1
Battery Empty Warning Function (Voltage) 2.65V±0.1
Leakage Current on Sleeping Mode (uA) Less than 250uA
Leakage Current on Standby Mode (mA) Less than 5mA
Maximum Current on Testing Mode (mA) Less than 9mA
Battery Lifetime Over 500 tests
LCD Display 41 x 37 (mm) TN LCD
Memory Capacity Over 500 Spk/Over 1,000 for Non Spk
Auto Turn-off 3 minutes
Date and Time Express Method Month-Day / Hour: Minutes / AM/PM
Daily Alarm Function 6 Sets
Meter Storage / Transportation Condition
(Temperature/Humidity)
Humidity: 15%~85%
Temperature 50ºF~104ºF (10 ºC~40 ºC)
Meter Operating Condition (Temperature/Humidity) Humidity: 15%~85%
Temperature -4ºF~140ºF (-20 ºC~60 ºC)
Measurement Units mg/dL USA / mg/dL (mmol/L) Int’l
Measurement Range 20~600 mg/dl (1.1~33.3 mml/L)
Test Time 5 seconds
Compatible with Test Strips (Model Nº) BMB-BA006A
Test Sample Volume 1 µ L depends on BMB-BA006A
PC Interface USB
The device has been certied to meet the electrical and safety requirements of:
IEC 61010-1, EN 61010-1, IEC 61010-2-101, EN 61010-2-101, IEC 61326-1, EN 61326-1, IEC 61326-2-6, EN 61326-2-6.