Section I- Product Introduction
11
Product Characteristics
User-friendly device management (all-in-one type/detachable type)
Decides if it will give a defibrillation shock to the patient or not after
analyzing the patient's ECG automatically.
(Defibrillation shock button should be pushed for delivering a
defibrillation shock.)
A non-rechargeable type lithium battery is used.
All-in-one type cartridge where electric pads and battery are
embedded is available. Also the detachable cartridge where electrode
pads and battery are separated is available.
Electrode pads can be used on both Adult and Pediatric patients.
(However, the attachment points are different.)
After analyzing ECG signals for 5 seconds, it stores them in the
internal memory (up to 1,000 cases during analysis only).
Stored data can be transmitted to a PC through the “NT-MPR” PC
management program.
Communication to PC is available with USB connection.
Ambient sound can be stored for up to four (4) cases and fifteen (15)
minutes for each case. (total memory: ~60 min.)
All operation procedures including CPR are guided by clear voice
prompts.
Self-diagnostics test function ensures device is in operational state.
Battery condition check in real time.
(In case of normal operational state, the pad attachment lamp
( ) flashes every 7 seconds.)