Section II - Product Use
29
Detailed Instructions (Pediatric)
1. Press power button ( ) for Pediatric mode.
Lamp state: Pediatric mode lamp ( ) will be
ON.
Ensure to check the electrode pad attachment
lamp ( ) is ON.
Pediatric refers to person under 8 years
(25kg).
2. Remove the patients top clothing.
If necessary, rip or cut off the clothing to bare the
persons chest
3. If there is any water on a patient's body before
attaching an electrode pad, dry the body with a
towel.
If the jacket was not taken off from the
patient, or water is not adequately removed
from the patients body, HeartPlus may not
analyze the patient’s state properly.
A defibrillation shock energy might not be
delivered to the patient’s heart.