Section IV Appendix
53
Electromagnetic Compatibility (EMC)
1) Electromagnetic Emission
Emission Test
Compliance
Electromagnetic environment Information
RF(Radio
Frequency)
CISPR11
Group 1
Class B
RF energy is only used for the internal
function of HeartPlus.
Therefore, self-RF emission level is very low;
in addition, it will not have effect on the
operation of electronic equipment near it.
This HeartPlus is available in household
building or general facilities where common
low voltage power is connected.
2) Electromagnetic immunity
Immunity test
IEC60601
Test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±15kV
contact
±15kV air
There are no special
requirements with respect to
electrostatic discharge.
Power
frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial/hospital environment.
There are no special requirements
for non-hospital environments.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.5GHz
20 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
HeartPlus, including cables, than is
absolutely necessary.