Section III - Guidelines
44
Cautions for Battery Pack
1. Do not heat battery pack or throw them into fire.
Heating the battery pack more than 100 degrees centigrade could
increase the internal pressure and electrolyte may boil causing resin parts
to melt, resulting in leakage, explosion or fire.
Also throwing the battery pack into fire can cause combustion of organic
solvents in the lithium-ion battery electrolyte, which is very dangerous.
2. Do not attempt to disassemble the battery pack.
If the battery pack is disassembled, it may short-circuit and cause metallic
lithium reaction with moisture in the air. The leaking electrolyte may
generate heat and catch fire, then the cartridge may emit smoke, burst or
ignite.
3. Battery pack storage recommendations.
In keeping the battery pack, your attention is required as follows :
1) When storing a battery pack out of the pouch, be sure that the
terminals of the battery pack is short-circuited.
2) Store it in a clean, dry, and cool place away from heat and metal
objects (before putting it into the device, or when outside of the device).
Avoid exposure to direct sunlight, high temperature or high humidity
when storing batteries, because these conditions may result in cartridge
degradation and leakage of liquid.
3) Do not lick the battery pack or put terminals on your lips.