Section II – Product Use
34
Battery Check
1. There are 2 types of warning for low battery.
1) In case of over about 80% use of the battery
Voice prompt: Before analyzing the patient, it tells you the battery
weakness with voice prompt “The battery is low, Change the battery”
The number of defibrillation shock available: maximum 10 times
Required time for defibrillation shock energy: 25~45 seconds
Action: You should replace it with a new battery
2) In case of over about 90% use of the battery charge
Attachment of electrode pad ( ), the electrocardiogram analysis
( ), and the cardiopulmonary resuscitation ( ) lamp are all
flashing.
Until the complete discharge of the battery “Change the battery”
prompt will sound 5 times at 30-minute intervals.
The number of defibrillation shocks available: not allowed
Action: Replace the battery immediately because you can use
HeartPlus™ only after replacing the battery in this situation. After
complete discharge of the battery, voice prompts will be stopped and
all the lamps will be off.
2. Voice prompts for equipment error
Voice prompt: “Equipment check is needed.”
Action: Replace the equipment immediately as it cannot be used.
Contact customer service center.