Section III - Guidelines
38
Check List - Prior to Use
1. HeartPlusis only for the sudden cardiac arrest patients.
Sudden cardiac arrest patients exhibit the following conditions :
1) Unconsciousness
2) No movement
3) No pulse
2. Before using the product, keep in mind the following :
1) Electric shocks are conductive to water, so move the patient to another place if
the surface is wet and remove any liquids off the skin of the patient prior to use.
2) Ensure that the equipment in an operational state, and not sounding error
warnings, before attempting use by checking the standby mode LED state.
* If the battery is normal, the electrode pad attachment lamp ( ) blinks every
7 seconds.
3. Take off patient's upper-body clothing and attach electrode pads.
1) Electrode pads should be adhered to the patient’s body in the same way as
illustrated on the picture on the pads themselves and on the AED’s main panel.
2) If the patients body is wet and contaminated, you should put electrode pads on
the body after using the supplied towel to clean and dry the patient. If for some
reason there is no towel, use any nearby cloth, such as the patient’s top clothing to
make sure the patient is clean and dry.
Failure to do so may result in incorrect ECG analysis.
4. While operating the device, pay close attention to the following :
1) The electrode pads should be apart from each other and/or metallic parts
touching the patient.
2) While HeartPlus is analyzing the patients state, the patient must not be
touched or moved, because this will result in an incorrect ECG analysis.
3) If on a bed or stretcher, avoid contact between patient’s exposed skin, such as