50
Defibrillation Waveform
Electric energy output system: Biphasic
[Fig1] Biphasic electric energy output waveform
[Table1] Adult mode energy discharge time according to the patient
impedance
Energy output parameter: Energy output time is automatically and
adjusted by the impedance of the patient.
(Error rate ± 10%)
Impedance(Ω)
Phase 1 duration (ms)
Phase 2 duration (ms)
Output energy(J)
25
2.8
2.7
181
50
4.4
4.4
181
75
6.1
6.1
178
100
7.9
7.9
179
125
9.7
9.7
175
150
11.5
11.5
178
175
13.3
13.3
175