Section IV Appendix
51
[Table2] Pediatric mode energy discharge time according to the
patient impedance
Energy output parameter: Energy output time is automatically and
adjusted by the impedance of the patient.
(Error rate ± 10%)
Impedance(Ω)
Phase 1 duration (ms)
Phase 2 duration (ms)
Output energy(J)
25
2.8
2.7
48.3
50
4.4
4.4
49.9
75
6.1
6.1
50.8
100
7.9
7.9
51.7
125
9.7
9.7
51.4
150
11.5
11.5
52.4
175
13.3
13.3
51.7