Section II Product Use
24
Detailed Instructions (Adult)
1. Press power button for adult ( ).
Lamp state: Adult mode lamp will be ON.
Adult refers to person older than 8 years (25kg).
After power switch ON, electrode pads
attachment lamp ( ) flickers until you attach
the electrode pads.
2. Remove the patients top clothing.
Note: If necessary, rip or cut off the clothing to bare
the persons chest.
3. If there is any water on the patients body before
attaching an electrode pads, dry the body with a
towel.
If the jacket was not taken off from the patient,
or water is not adequately removed from the
patients body, HeartPlus may not analyze the
patient’s state properly.
A defibrillation shock energy might not be
delivered to the patient’s heart.