Discharge Current Waveform: BTE (Biphasic Truncated
Exponential)
Discharge Time: Automatically adjusted energy for
patient impedance
Output Energy:
- 180J (for adult, load impedance of 50)
- 50J (for infant, load impedance of 50)
Patient Insulation: BF Type
Charging period: Less than 13 seconds after pads
attachment.
Operation Mode: AED (automated external
defibrillator), which can analyze ventricular fibrillation
Change Method to Pediatric Mode: Power on device
in Pediatric Mode (small yellow button)
Composition: Disposable Electrode
pads, Battery
Frequency of Use: One time
Packing Method: Sealing Packaging
Size: 150(W) X 110(L) X 33(H)mm
Warranty: 2 years (Sealed storage at
room temperature)
Frequency of Use : Single Use
Size: 125mm X 95mm
Area: 76.63㎠ (PE Foam, Hydrogel)
Wire Length: 1.2m
Attachment point: Showed in picture on