Section III - Guidelines
42
Only all-in-one cartridges that are supplied by
NANOOMTECH should be used for our
product.mightm
Do not open all-in-one cartridge or detachable
electrode pads pouch if you are not going to use
them right away.
As soon as you open all-in-one cartridge or
Detachable electrode pads, adhesive element of
electrode pads starts to dry regardless of expiry
date and it cannot be used.
4) Perform self-test referring to self-test method specified in Chapter 2
Use of Product. After finishing self-test, when state light flickers in green
color, it means equipment is operated normally. If you have any problem,
contact our customer service center.
5) After you keep the product in designated place, you need to be
prepared to use it when necessary.
Maintaining Product Cleanliness and Its Cleaning
You can clean the exterior of the product using moistened soft cloth by
soapy water, light chlorine bleach (2 teaspoons per 1 liter water) or
cleaner which is made of ammonia.
Do not soak the product or accessories in liquid or
detergent.
Keep the product from entering of any liquid into it.
Do not use the detergent containing abrasive
ingredients.
When product is submerged into the water, contact
our customer service center immediately.
Do not sterilize the product or accessories.