Micro-Ohmmeter Model 5600
Operating Procedure
Before performing the resistance test, verify that the sample under test is
not energized.
Connections and Reading
Clean all surfaces before connecting test leads. Verify a solid connection
between test leads and the sample. Set the On/Off button to the ON
position and select the range that provides the most stable reading. If the
anticipated resistance is not known, begin with the highest range (200Ω)
and successively lower the range selection until adequate resolution is
achieved. The range selection may be changed while the instrument
remains energized.
Read the displayed resistance value directly. The use of a Kelvin
connection eliminates test lead resistance, as shown in Figure 5.
UI
C
2
P2
C1
P1
Rx
MICRO-OHM
!
F1
- I
®
INSTRUMENTS
Figure 5
- 12 -