EasyManuals Logo

Aethlabs microAeth MA200 User Manual

Aethlabs microAeth MA200
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
74 microAeth
®
MA Series MA200, MA300, MA350 Operating Manual microAeth
®
MA Series MA200, MA300, MA350 Operating Manual 75
AethLabsAethLabs
7.3.2.2. Flow Calibration Table Example
Below is an example of a ow calibration table that will be displayed using the Display All Settings or
Display FlowCal Info menu items. The values in the ow calibration table can be analyzed and veried in
order to determine if another ow calibration is required. If multiple ow calibrations have been performed
and still there are issues, please contact AethLabs.
FlowCal Table
Device ID: MA200-0011
Flow Meter Serial Number: 146654
Timestamp: 2018-11-20T17:17:20.00
Battery: 4.245V
Sample temp: 22.51C
Sample RH: 37.63%
Internal pressure: 101536Pa
SingleSpot Pump Start: 100
SingleSpot Pump Stop: 12
SingleSpot Flow Min: 19
SingleSpot Flow Max: 281
DualSpot Pump Start: 5
DualSpot Pump Stop: 1
DualSpot Flow Min: 21
DualSpot Flow Max: 299
-----------SingleSpot Mode----------- -----------DualSpot Mode------------
# TargetS SFlowT SFlow1 PumpS SSD TargetD DFlowT DFlow1 PumpD DSD S1/S2
0
0.00 124.99 124.03 0 0.00 0.00 124.90 123.90 0 0.01 0.00
1
43.93 180.30 179.20 69 2.04 45.75 181.79 164.51 49 2.46 1.41
2
58.82 197.69 197.76 88 1.65 58.74 197.48 175.92 65 1.82 1.42
3
74.38 215.58 216.97 112 1.21 74.90 215.49 190.49 87 1.48 1.41
4
89.79 231.66 235.80 140 0.94 90.07 231.28 203.94 111 1.21 1.41
5
104.83 245.98 253.85 170 0.78 105.09 245.39 217.09 139 0.99 1.41
6
120.17 259.53 271.43 203 0.60 120.35 258.28 229.99 169 0.86 1.41
7
135.40 272.50 288.69 238 0.53 135.78 271.66 243.48 201 0.68 1.41
8
149.71 283.43 304.11 273 0.45 150.28 282.74 255.69 233 0.60 1.41
9
164.67 293.83 319.65 315 0.37 164.56 293.13 267.65 267 0.51 1.40
10
179.39 303.29 334.45 360 0.31 179.79 303.28 280.27 306 0.42 1.40
11
194.06 310.21 348.08 409 0.25 194.67 311.80 292.35 348 0.37 1.40
12
209.38 319.61 362.21 469 0.22 209.56 321.00 303.94 396 0.33 1.40
13
223.92 327.67 375.51 535 0.19 224.24 329.85 315.56 450 0.26 1.39
14
238.91 334.19 388.55 617 0.17 239.08 338.19 326.65 512 0.22 1.39
15
253.92 340.43 401.32 728 0.14 253.44 343.88 337.23 584 0.24 1.38
16
269.59 344.87 413.57 843 0.21 268.32 349.48 348.23 680 0.20 1.38
17
0.00 0.00 0.00 0 0.00 283.32 357.21 359.09 789 0.14 1.37
18
0.00 0.00 0.00 0 0.00 0.00 0.00 0.00 0 0.00 0.00
19
0.00 0.00 0.00 0 0.00 0.00 0.00 0.00 0 0.00 0.00
20
0.00 0.00 0.00 0 0.00 0.00 0.00 0.00 0 0.00 0.00
21
0.00 0.00 0.00 0 0.00 0.00 0.00 0.00 0 0.00 0.00
FlowCal Results: OK 35/35
7.3.3. Test Flow Procedure
The on-board user interface on the front of the microAeth can be used to test ow setpoints and see what
the internal owmeters are reading in DualSpot or SingleSpot mode. Please note that the external owmeter
of the microAeth MA Series Flow Calibration Kit or other low backpressure external owmeter is not needed
to use this feature but is needed to verify the internal measured ow readings against an external standard.
1) If checking the internal measured ow of the microAeth to an external standard, connect the outlet port
of the external owmeter to the inlet port of the microAeth.
2) Make sure that all connections are tight and sealed so that no leaks are present.
3) Use the left and right buttons of the microAeth to scroll through the top level menu options to the 'Test
Flow' option.
4) Press the center button of the microAeth to select the ‘Test Flow’ option.
5) Use the left and right buttons of the microAeth to scroll to the desired target ow of 50, 75, 100, 125, or
150 ml/min. The screen will display the following depending if the microAeth is in SingleSpot or DualSpot
mode:
SingleSpot Mode Test Flow Display DualSpot Mode Test Flow Display
SS Target xxx.xx DS Target xxx.xx
SS Measured xxx.xx DS Measured xxx.xx
NOTE: To use the test ow feature in the desired SingleSpot or DualSpot mode, this sampling mode setting
must rst be changed using the microAeth Manager software connected to the microAeth by USB.
6) Use the left and right buttons to change the target ow at any time. Allow the microAeth a few seconds
to adjust pump speed and for the ow to become stable.
7) The measured ow can be viewed over time as there will be uctuations. The measured ow can also
be compared against the external owmeter of the MA Series Flow calibration Kit that has its outlet port
connected to the inlet port of the microAeth. Only compatible low backpressure external owmeters should
be used.
8) Press the center button to exit the test ow feature.

Table of Contents

Other manuals for Aethlabs microAeth MA200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Aethlabs microAeth MA200 and is the answer not in the manual?

Aethlabs microAeth MA200 Specifications

General IconGeneral
BrandAethlabs
ModelmicroAeth MA200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals