4454C EN 2013-05-14Safety precautions
CR 35-X DIGITIZER
The classification of this product according to the medical electrical
equipment standard IEC 60601-1 requires installation outside the patient
vicinity. For definition of patient vicinity see dimensions below.
R = 1.5 m/4.9 feet (EN 60601-1) or 1.83 m/6 feet (UL 60601-1).
h = 2.5 m/8.2 feet (EN 60601-1) or 2.29 m/7.5 feet (UL 60601-1).
Markings and labels
Always take into account the markings and labels provided on the inside and
outside of the machine. A brief overview of these markings and labels and
their meaning is given below.
WARNING: To avoid risk of electric shock, this equipment must only
be connected to a supply mains with protective earth.
Patient vicinity
Patient vicinity
R
h