HDR-DAD Solution 45
Using the HDR-DAD Solution
4
HDR-DAD Solution - Data Analysis
HDR-DAD Solution - Data Analysis
The HDR-DAD solution offers the following sections for Data Analysis:
Three separate signal groups for HDR, DAD
1
(LNG = long cell), DAD
2
(SHT = short cell)
The user can observe 24 signals 3 (HDR, DAD
1
, DAD
2
) x 8 signals (A-H)
Spectra information
NOTE
Display of Delay Time in the data analysis software
The flow from DAD
1
- capillary - DAD
2
causes a Delay Time. The HDR-DAD solution
calculates the chromatogram from DAD
1
and DAD
2
signals. Therefore a
HDR-chromatogram always starts at earliest with the signal of DAD
2
.
In spite of this, ChemStation and EZChrom display the delay time differential.
ChemStation:
The HDR-DAD chromatogram shows a short gap at the beginning of the signal. This
reflects the fact that the combined HDR signal can only start as it is delayed by the
delay time between the detectors. Whereas the signals of the single detectors start
immediately at t = 0 min.
See Figure 24 on page 46.
EZChrom:
The HDR-DAD chromatograms always start at t = 0 min and shows a short gap at the
end of the signal.
See Figure 25 on page 46.

Other manuals for Agilent Technologies 1200 Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1200 Infinity Series and is the answer not in the manual?

Agilent Technologies 1200 Infinity Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1200 Infinity Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals