EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>33522A

Agilent Technologies 33522A User Manual

Agilent Technologies 33522A
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
136
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Modulating Waveform Frequency
The waveform generator will accept an internal or external modulation
source for PWM. The external modulation input has a -3dB bandwidth of
100 kHz.
Modulating frequency (internal source): 1 Hz to 30 MHz.
The default
is 10 Hz
.
Front-Panel Operation
: After enabling PM, press the
PWM Freq
softkey.
Remote Interface Operation
:
[SOURce[1|2]:]
PWM:INTernal:FREQuency
{<
frequency
>|MINimum|MAXimum}
Phase Deviation
The phase deviation setting represents the peak variation in phase of the
modulated waveform from the carrier waveform. The phase deviation
can be set from 0 to 360 degrees. The default is 180 degrees.
Front-Panel Operation
: After enabling PWM, press
Phase Dev
. Then
use the knob or numeric keypad to enter the desired deviation.
Remote Interface Operation:
PWM:DEViation {<deviation_in_degrees>|MINimum|MAXimum}

Table of Contents

Other manuals for Agilent Technologies 33522A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33522A and is the answer not in the manual?

Agilent Technologies 33522A Specifications

General IconGeneral
BrandAgilent Technologies
Model33522A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals