Installation
18 Agilent 4107 Nitrogen Generator User’s Guide
3 While still attached to the rails, use an overhead or portable hand
crane and slings, thread the slings between the wheels and the
substructure, both front and back and then lift the nitrogen
generator.
4 Remove the pallet from underneath and then lower the nitrogen
generator to the floor.
5 Once the instrument is completely off of the pallet, unscrew the
three screws attaching the rail to the nitrogen generator.
6 The nitrogen generator can then be wheeled into position.
Position the nitrogen generator to allow easy access to the power
button located on the front of the instrument.
Figure 3. Nitrogen generator tubing and power connections
Air supply hose
Power input
Nitrogen output
Filtered air
supply input
Fitting to air
compressor