4. Operation
Before Operation 21
Turning on the Nitrogen Generator 21
Selecting the Nitrogen Generator 22
Purging the Nitrogen Generator 23
Turning Off the Nitrogen Generator 23
Depressurizing the Nitrogen Generator 24
Before Operation
Before starting the nitrogen generator, carefully read the ‘Safety
Practices and Hazards’ section at the front of this guide and ensure
that the laboratory is set up according to the details specified in the
Agilent MP-AES Site Preparation Guide.
In operation, the nitrogen generator cycles approximately every 40
seconds. At the end of each cycle the oxygen enriched air is vented to
the room from a vent in the bottom of the nitrogen generator. A noise
may be heard as this gas is released.
Turning on the Nitrogen Generator
To turn on the nitrogen generator:
1 Ensure that the gas filters are installed correctly and do not need
replacing.
2 Ensure the air and nitrogen hoses are connected to the nitrogen
generator and that the air supply is turned on.
3 Check the pressure indicated on the pressure gauge of the filter
regulator unit on the right side of the nitrogen generator meets
the specification.