EasyManuals Logo

Agilent Technologies 4156C User Manual

Agilent Technologies 4156C
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4 2-69
Driver Function Reference
hpe5250a_self_test
the R, L, and C value, measure R, L, and C of the cable using the 4284A, and divide
them by cable length (in m). Compensation data must be the value for 1 m length.
Do not change the data format in the file.
hpe5250a_self_test
This function causes the instrument to perform a self-test and returns the result of
that self-test. This is used to verify that an instrument is operating properly. A failure
may indicate a potential hardware problem.
Syntax ViStatus _VI_FUNC hpe5250a_self_test(ViSession vi, ViPInt16 test_result,
ViChar_VI_FAR test_message[ ]);
Parameters vi Instrument handle returned from hpe5250a_init( ).
test_result Numeric result from self-test operation. 0: No error.
test_message[ ] Self-test status message. This is limited to 256 characters.
hpe5250a_testClear
This function clears the test result for the specified relay card or the front panel or
the controller.
Syntax ViStatus _VI_FUNC hpe5250a_testClear(ViSession vi, ViInt16 framecard_clear);
Parameters vi Instrument handle returned from hpe5250a_init( ).
framecard_clear Test result to be cleared. 0 (test result of all test), 1 (card 1 relay
test result), 2 (card 2 relay test result), 3 (card 3 relay test
result), 4 (card 4 relay test result), 5 (relay test result of all
card), 6 (front panel test result), or 7 (controller test result).
# E5250A C Compensation coefficient data table
#
# CAUTION : Do not add or delete "REVISION" line and "DATAxx" line.
# Change the value for R,L,C of DATA05,06,07 or 08.
#
REVISION A.03.00
# R [ohm] L [H] C [F]
DATA00 74.65E-3 140.00E-9 58.44E-12 # Frame Path 1
DATA01 75.41E-3 90.00E-9 67.13E-12 # Frame Path 2
DATA02 231.41E-3 450.00E-9 178.85E-12 # Card Path High
DATA03 177.56E-3 390.00E-9 135.45E-12 # Card Path Low
DATA04 100.70E-3 400.00E-9 80.00E-12 # Triax Cable [/m]
DATA05 100.70E-3 400.00E-9 80.00E-12 # User Triax Cbl H [/m]
DATA06 100.70E-3 400.00E-9 80.00E-12 # User Triax Cbl L [/m]
DATA07 114.00E-3 544.00E-9 130.00E-12 # User Coax Cbl H [/m]
DATA08 114.00E-3 544.00E-9 130.00E-12 # User Coax Cbl L [/m]
DATA09 0.00E-3 0.00E-9 1.20E-12 # Stray Capacitance
# END of Data

Table of Contents

Other manuals for Agilent Technologies 4156C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4156C and is the answer not in the manual?

Agilent Technologies 4156C Specifications

General IconGeneral
BrandAgilent Technologies
Model4156C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals