Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4 2-55
Driver Function Reference
hpe5250a_closeList
Parameters vi Instrument handle returned from hpe5250a_init( ).
close_card Card number. 1 (card 1), 2 (card 2), 3 (card 3), or 4 (card 4) in
the normal configuration mode, or 0 (all card in the automatic
configuration mode). For the configuration mode, see
hpe5250a_func.
closechan_list[ ] Channels closed of the specified card.
hpe5250a_closeList
This function will connect the input ports to the output ports specified by the
channel list.
The parameter ’closechan_list’ is an array of integers with each integer representing
one channel. The last number of the ’closechan_list’ should be “0” (numeric zero) to
identify the end of the list. The maximum number of channels that can be specified
by the list is 100.
Syntax ViStatus _VI_FUNC hpe5250a_closeList(ViSession vi,
ViInt32_VI_FAR closechan_list[ ]);
Parameters vi Instrument handle returned from hpe5250a_init( ).
closechan_list[ ] Channel numbers to connect. 5 digits integer. ABCDE. where
A: card number, BC: input port number, DE: output port
number. Top zero(s) can be ignored. For example, if A=0,
BC=01, and DE=01, channel number should be 101 instead of
00101.
hpe5250a_closeList_Q
This function will query the instrument for the channels closed given in the
’closechan_list’.
The parameter ’closechan_list’ is an array of integers with each integer representing
one channel. The last number of the ’closechan_list’ should be “0” (numeric zero) to
identify the end of the list. The maximum number of channels that can be specified
by the list is 100.
The ’close_status’ parameter is an array of integers containing the return values of
the query. The ’close_status’ array returned will correspond one to one with
’closechan_list’ parameter.