15.
INPUT A
TTENUATOR
SWITCHING
UNCERT
AINTY
TEST
Input
R
Input
A
tten
uation
Minim
um
Limit
T
est
Result
Maxim
um
Limit
Measurement
Uncertain
t
y
0
dB
0
1
dB
1
dB
6
0.017
dB
20
dB
0
1
dB
1
dB
6
0.016
dB
30
dB
0
1
dB
1
dB
6
0.016
dB
40
dB
0
1.5
dB
1.5
dB
6
0.016
dB
50 dB
0
1.5
dB
1.5 dB
6
0.024
dB
Input
A
Input
A
tten
uation
Minim
um
Limit
T
est
Result
Maxim
um
Limit
Measuremen
t
Uncertain
t
y
0dB
0
1
dB
1dB
6
0.017
dB
20
dB
0
1
dB
1
dB
6
0.016
dB
30
dB
0
1dB
1
dB
6
0.016 dB
40
dB
0
1.5
dB
1.5
dB
6
0.016
dB
50
dB
0
1.5
dB
1.5
dB
6
0.024
dB
Input
B
Input
A
ttenuation
Minim
um Limit
Test
Result
Maxim
um
Limit
Measuremen
t
Uncertain
ty
0
dB
0
1
dB
1
dB
6
0.017
dB
20
dB
0
1
dB
1
dB
6
0.016
dB
30
dB
0
1
dB
1
dB
6
0.016
dB
40
dB
0
1.5
dB
1.5
dB
6
0.016
dB
50 dB
0
1.5 dB
1.5 dB
6
0.024 dB