16.
NOISE SIDEBANDS
TEST
Input
R
F
requency
Oset
from
Carrier
T
est
Result
T
est
Limit
Measuremen
t
Uncertain
t
y
10
MHz
0
1
kHz
<
0
97
dBc/Hz
6
2.14
dB
+1
kHz
<
0
97 dBc/Hz
6
2.14
dB
0
10
kHz
<
0
97
dBc/Hz
6
0.98
dB
+10
kHz
<
0
97
dBc/Hz
6
0.98
dB
0
100
kHz
<
0
110 dBc/Hz
6
2.14
dB
+100
kHz
<
0
110
dBc/Hz
6
2.14
dB
0
1
MHz
<
0
110
dBc/Hz
6
0.98
dB
+1
MHz
<
0
110
dBc/Hz
6
0.98
dB
100
MHz
0
1kHz
<
0
97
dBc/Hz
6
2.14 dB
+1
kHz
<
0
97
dBc/Hz
6
2.14
dB
0
10
kHz
<
0
97
dBc/Hz
6
0.98
dB
+10
kHz
<
0
97
dBc/Hz
6
0.98
dB
0
100
kHz
<
0
110
dBc/Hz
6
2.14
dB
+100
kHz
<
0
110
dBc/Hz
6
2.14
dB
0
1
MHz
<
0
110
dBc/Hz
6
0.98
dB
+1
MHz
<
0
110
dBc/Hz
6
0.98
dB
500
MHz
0
1
kHz
<
0
97
dBc/Hz
6
2.14
dB
+1
kHz
<
0
97
dBc/Hz
6
2.14
dB
0
10 kHz
<
0
97
dBc/Hz
6
0.98
dB
+10
kHz
<
0
97
dBc/Hz
6
0.98
dB
0
100
kHz
<
0
110
dBc/Hz
6
2.14
dB
+100
kHz
<
0
110
dBc/Hz
6
2.14
dB
0
1
MHz
<
0
110
dBc/Hz
6
0.98
dB
+1 MHz
<
0
110 dBc/Hz
6
0.98 dB
Performance Test Record 4-17