Installation
Agilent 5100 and 5110 ICP-OES User’s Guide 33
ICP-OES torches
Many of the ICP-OES torches are listed below, each with unique
properties to suit a wide range of applications. Additional torches
may be available. See the Agilent website for more information.
Parameter
Dual View Torches
Standard
torch
Demountable
torch
Inert demountable
torch
High solids
demountable
torch
Semi-volatile
organics
demountable torch
Construction One piece
Two piece
demountable
Two piece
demountable
Two piece
demountable
Two piece
demountable
Injector diameter
(mm) 1.8 1.8 1.8 2.4 1.4
Injector material Quartz Quartz Ceramic Quartz Quartz
Outer tube length
Medium
with slot
Medium with
slot
Medium with slot Medium with slot Medium with slot
Outer tube material Quartz Quartz Quartz Quartz Quartz
Intermediate tube Quartz tulip Quartz tulip Quartz tulip Quartz tulip Quartz tulip
Recommended
application
Most
sample
types
Most sample
types
Hydrofluoric (HF)
acid digests
Samples with
high total
dissolved solids
Semi-volatile
organic solvents
e.g., ketones,
kerosene, Jet-A1
Parameter
Radial View Torches
Standard torch Demountable torch
Inert demountable
torch
Volatile organics
demountable torch
Construction One piece
Two piece
demountable
Two piece
demountable
Two piece
demountable
Injector diameter (mm) 1.4 1.4 1.8 0.8
Injector material Quartz Quartz Ceramic Quartz
Outer tube length Short (no slot) Short (no slot) Short (no slot) Short (no slot)
Outer tube material Quartz Quartz Quartz Quartz
Intermediate tube Quartz tulip Quartz tulip Quartz tulip Quartz tulip
Recommended
application
Most sample types Most sample types
Hydrofluoric (HF)
acid digests
Volatile organic
solvents e.g.,
gasoline