Operation
46 Agilent 5100 and 5110 ICP-OES User’s Guide
Preparing for Analysis
To prepare for analysis:
1 Click the Plasma button in the ICP Expert software.
Alternatively, press F5 or choose Plasma on from the arrow
under the Plasma button.
NOTE
The plasma ignition sequence will take up to 60 seconds to complete. If the
plasma fails to ignite, refer to the Troubleshooting section in the Help for further
information.
NOTE
For optimum performance and stability, for the ICP-OES instrument a warm up
time of 20 minutes is recommended after igniting the plasma.
If using wavelengths below 189 nm: purging the polychromator can take up to
several hours. The polychromator thermal stabilization can take up to several
hours from long term shutdown.
2 Ensure that the peristaltic pump is correctly set up (refer to the
Peristaltic Pump section of the ICP-OES Familiarization DVD or
the ICP Expert Help). If you have not already done so, adjust the
pressure bars on the peristaltic pump for even sample flow and
drainage.
3 Place the sample tubing from the peristaltic pump into the rinse
solution and the drain tubing into the drainage vessel.
4 Click the Pump button in the ICP Expert software and choose
Normal (15 rpm) from the arrow under the Pump button. The
pump will be initialized and the solution will begin aspirating.
Performing a Detector and Instrument Calibration
See Page 30 for instructions.

Table of Contents

Other manuals for Agilent Technologies 5110 ICP-OES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5110 ICP-OES and is the answer not in the manual?

Agilent Technologies 5110 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5110 ICP-OES
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals