5 Operating in CI Mode
120 5977B Series MSD Operating Manual
General Guidelines
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
Always verify the MSD is performing well in EI mode before switching to CI.
(See “To Verify EI System Performance” on page 108.)
Ensure the CI source and GC/MSD interface tip seal are installed.
Ensure the reagent gas plumbing has no air leaks. This is determined in PCI
mode, checking for m/z 32 after the methane pretune.
Ensure the reagent gas inlet line is equipped with gas purifiers (not
applicable for ammonia).