General Maintenance 4
5975/5977 Series MSD Operation Manual 105
To Clean the EI Ion Source
Materials needed
• Abrasive paper (5061-5896)
• Alumina abrasive powder (8660-0791)
• Aluminum foil, clean
• Cloths, clean (05980-60051)
• Cotton swabs (5080-5400)
• Glass beakers, 500 mL
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Solvents
• Acetone, reagent grade
• Methanol, reagent grade
• Methylene chloride, reagent grade
• Ultrasonic bath
Preparation
1 Disassemble the ion source. See “To Disassemble the Standard or Inert EI
Ion Source” on page 99 or “To Disassemble the Extractor EI Ion Source” on
page 102.
2 Collect the following parts to be cleaned for a standard or inert EI ion
source: (See Figure 15 on page 107.)
• Repeller
• Interface socket
• Source body
• Drawout plate
• Drawout cylinder
• Ion focus lens
• Entrance lens