Intuvo 9000 GC Column Maintenance 3
5977B Series MSD Operating Manual 69
20 Tighten the Guard Chip compression bolt using the provided torque driver
until you hear one click.
21 Install the inlet cover.
22 Install the bus door on its hinges and close it.
23 Close the GC front door.
24 If removed, install the ALS injector.
Use the Intuvo torque driver to tighten the compression bolt until you hear one click.
Over-tightening can damage the flow path, strip the fittings, and cause leaks.