EasyManua.ls Logo

Agilent Technologies 5977 MSD

Agilent Technologies 5977 MSD
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction 1
5975/5977 Series MSD Operation Manual 21
Equipment precaution
You MUST make sure the front side-plate thumbscrew is fastened finger-tight.
Do not overtighten the thumbscrew; it can cause air leaks.
You must remove the plastic cover over the glass window on the front of a
5977B Series MSD. In the unlikely event of an explosion, this cover may
dislodge.
General laboratory precautions
Avoid leaks in the carrier gas lines. Use leak-checking equipment to
periodically check for hydrogen leaks.
Eliminate from your laboratory as many ignition sources as possible (open
flames, devices that can spark, sources of static electricity, etc.).
Do not allow hydrogen from a high pressure cylinder to vent directly to
atmosphere (danger of self-ignition).
Use a hydrogen generator instead of bottled hydrogen.
Operating precautions
Turn off the hydrogen at its source every time you shut down the GC or
MSD.
Turn off the hydrogen at its source every time you vent the MSD (do not
heat the capillary column without carrier gas flow).
Turn off the hydrogen at its source every time shutoff valves in an MSD are
closed (do not heat the capillary column without carrier gas flow).
Turn off the hydrogen at its source if a power failure occurs.
If a power failure occurs while the GC/MSD system is unattended, even if
the system has restarted by itself:
1 Immediately turn off the hydrogen at its source.
2 Turn off the GC.
3 Turn off the MSD and allow it to cool for 1 hour.
WARNING
Failure to secure your MSD as described above greatly increases the chance of
personal injury in the event of an explosion.

Table of Contents

Related product manuals