58 5977B Series MSD Operating Manual
3 Intuvo 9000 GC Column Maintenance
Columns
Many types of Intuvo 9000 GC columns can be used with the MS, but there are
some restrictions.
During tuning or data acquisition, the rate of column flow into the MS should
not exceed the maximum recommended flow. Therefore, there are limits to
column length and flow. Exceeding recommended flow will result in
degradation of mass spectral and sensitivity performance.
Remember that column flows vary greatly with temperature, requiring
measurements of the actual flow to be taken. Use Table 6 to determine an
acceptable column flow.
Table 6 Gas flows
Feature Gas flows
High-vacuum pump Turbo
Optimal He column gas flow, mL/min (carrier gas) 1 to 2
Maximum recommended gas flow, mL/min
*
* Total gas flow into the MS = column flow
4 to 6
Maximum gas flow, mL/min
Expect degradation of spectral performance and sensitivity
6.5
Maximum column id 0.53 mm (30 m length)

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals