60 5977B Series MSD Operating Manual
3 Intuvo 9000 GC Column Maintenance
Handling the Intuvo 9000GC Column and Bus Components
The Agilent Intuvo 9000 Gas Chromatograph (Intuvo 9000 GC) does not use
traditional ferrules and nuts for most column and flow path seals. In a
traditional gas chromatography connection, the seal is made by deforming a
soft ferrule around the periphery of a column or tube, with a second seal made
between the ferrule and the fitting. Instead, the Intuvo 9000 GC click and run
connections use a sealing system based on contact between flat surfaces.
Compared to traditional ferrule seals, these connections are leak-free and easy
to make.
When making these seals, follow a few simple guidelines:
Do not touch the click and run sealing surfaces with bare skin or dirty
gloves. Skin oils and dirt can contaminate the flow path surfaces.
Use only the provided Intuvo 9000 GC torque driver to tighten Intuvo
compression bolts.
Avoid scratching or deforming click and run sealing surfaces.
If you need to clean a sealing surface, use clean, compressed air.
Use a new gasket each time you install a column or Intuvo chip.

Table of Contents

Other manuals for Agilent Technologies 5977 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977 MSD and is the answer not in the manual?

Agilent Technologies 5977 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5977 MSD
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals