EasyManua.ls Logo

Agilent Technologies 6460 - Page 18

Agilent Technologies 6460
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Exercise 1 – Develop an acquisition method
Task 3. Find optimum fragmentor voltage for maximum response
18 Agilent 6400 Series Triple Quad LC/MS Familiarization Guide
d If necessary, click Define
Chromatograms in the Chromatogram
section of the Method Explorer.
e To delete the BPC chromatogram, click
Delete in the Method Editor window.
f Select EIC for the Chromatogram
Definition Type,
g In the MS Chromatogram tab, make
sure MS Level is set to All and Scans
is set to All scan types.
h Clear the Do cycle sum check box.
i Type
271 as the m/z value.
j Click Add.
k Repeat steps i and j for the other
precursor ions,
279, 285 and
311.
l Click Method > Save As. The system
opens the Save As dialog box
m Save the method as iiiExercise 1.m.
n Click Save.
The default Method Editor list
selection after installation is
Integrate (MS).
You can also select Define
Chromatograms from the Method
Items list in the Method Editor
window.
When you are defining
chromatograms, instead of deleting
the Defined chromatogram, you can
select EIC. Then, you enter the m/z
value and click the Change button.
Steps Detailed Instructions Comments

Related product manuals